MR Pediatric Thoracic Spine WO Protocol
Scan Notes
Last updated: 11/3/2020
Charge as: TSP WO
Scanner preference: 1.5T or 3T
Coil: 16ch NV with Anterior Element (do not use top-off in MR1 if possible)
0 days-2wks = 3mm skip 0mm / Sag FOV= 15cm / Axial FOV=12cm
2wks - 2 yrs = 3mm skip 0.3mm / Sag FOV= 18cm / Axial FOV=12cm
2yrs - 9 yrs = 3mm skip 0.3mm / Sag FOV= 18cm / Axial FOV=15cm
Plane | Weighting | Mode | Slice | Gap | FAT SAT | FOV | Notes |
---|---|---|---|---|---|---|---|
SAG | T2 DRIVE | TSE | 3mm | 0mm - 0.3mm | None | 15-18cm | Gap and FOV per patient age. Lateral Sides of Vertebral Bodies |
SAG | T1 | TSE | 3mm | 0mm - 0.3mm | None | 15-18cm | Gap and FOV per patient age. Lateral Sides of Vertebral Bodies |
SAG | STIR | TSE | 3mm | 0mm - 0.3mm | STIR | 15-18cm | Gap and FOV per patient age. Lateral Sides of Vertebral Bodies |
AXIAL | T2 | TSE | 3mm | 0mm - 0.3mm | None | 12-15cm | Gap and FOV per patient age. Mid C7 to Mid L1. |