MR Pediatric Epilepsy/Seizure Brain WWO Neuro Protocol - Ingenia

Scan Notes: Above 9 years use Adult Protocol

Last updated:3/21/2019
Charge as: Brain WWO
Scanner preference: Ingenia Scanners Only: MR4, DMR2, BEAVERTON. 3T only. If patient has an implant unsafe for 3T, OK to scan on MR2 1.5T Ingenia. 
Coil: Head

Plane Weighting Mode Slice Gap FAT SAT FOV Notes
3D AXIAL SWI T1 FFE 2mm -1mm None 23cm Angle Parallel to temporal lobe. Ok to add slices.
OBL AXIAL T2 TSE 3mm 1mm None 23cm Parallel to Temporal Lobe
AXIAL DWI 2mm Voxel SE EPI 3mm 0.3mm SPIR 23cm Angle to Corpus- Skull Base to Vertex
OBL COR T2 STIR TSE 2mm 0.2mm STIR 23cm Whole brain, perpendicular to temporal lobe
3D OBL COR Flair BrainView TSE IR - 0mm None 25cm Whole brain, perpendicular to temporal lobe. Ok to add slices.
SAG 3D T1 MPRAGE 3D TFE 1mm 1mm None 23cm Extend slice coverage 1-2 mm beyond the skin margin, not necessarily to cover the ears to ensure adequate signal on the AX and COR MPRs. AX MPR: 1.3 x 1.3OBL COR MPR: 1 x 1 perpendicular to temporal lobe

Contrast injection

Plane Weighting Mode Slice Gap FAT SAT FOV Notes
AXIAL POST T1 TSE 3mm 1mm None 23cm Angle to Corpus- Skull Base to Vertex
COR POST T1 FAT SAT TSE 3mm 1mm SPIR 23m Angle to Corpus. Frontal through Occipital Bone
MR Protocol Epilepsy Seizure Brain for Ingenia in Radiology