MR Brain Tumor Follow Up WWO + Perfusion Neuro Protocol
Scan notes:
Adjust FOV and slices to best fit the patient's age and size:
- 0-2 years: 3mm slices / 1mm gap /small FOV
- 2-9 years: 4mm slices / 1mm gap / smaller FOV
- Over 9 years and adults: 4mm slices/1mm gap / larger FOV
- Send DSC source images to PACS, Intellispace and Archive
- Send Post Ax T1 to Intellispace
- Post process DSC
- Archive all sequences
POWER INJECTION
DSC: Injection Rate: 4-5mL/sec. Start injection 8-10 seconds after beginning the sequence. For peds patients with smaller IVs, consult with Rad and peds RN or Anesthesiologist to determine a safe injection rate.
Last updated: 1/7/2020
Charge as: Brain WWO
Scanner preference: OK to run this protocol if not scheduled on Ingenia scanner. Preferred on Ingenia scanners (MR4, MR 2, DMR2, BEAVERTON)
Coil: Head
Plane | Weighting | Mode | Slice | Gap | FAT SAT | FOV | Notes |
---|---|---|---|---|---|---|---|
SAG | T1 | TSE | 4mm | 1mm | None | 23cm | Scalp to Scalp |
AXIAL | T1 | TSE | 4mm | 1mm | None | 23cm | Angle to Corpus- Skull Base to Vertex |
AXIAL | T2 | TSE | 4mm | 1mm | None | 23cm | Angle to Corpus- Skull Base to Vertex |
COR | Flair | TSE | 4mm | 1mm | None | 23cm | Frontal through Occipital Bone |
AXIAL | T2* GRE | GRE | 4mm | 1mm | None | 23cm | Angle to Corpus- Skull Base to Vertex |
AXIAL | DWI 2mm Voxel | SE EPI | 3mm | 0.3mm | SPIR | 23cm | Angle to Corpus- Skull Base to Vertex |
Contrast injection
Plane | Weighting | Mode | Slice | Gap | FAT SAT | FOV | Notes |
---|---|---|---|---|---|---|---|
DSC | Run on both 1.5T & 3T | - | - | - | - | - | Inject 8-10 seconds after dynamics begin. Cover lesion and as much brain tissue as possible. Do not add slices. |
AXIAL | T1 | TSE | 4mm | 1mm | None | 23cm | Same as pre |
COR | T1 FAT SAT | TSE | 4mm | 1mm | SPIR | 23cm | Same as pre |
SAG | T1 | TSE | 4mm | 1mm | None | 23cm | Same as pre |