MR Trauma CSP WO Protocol

Last updated: 10/16/2020
Charge as: CSP WO
Scanner preference: 1.5T or 3T
Coil: 16ch NV with Anterior Element (do not use top-off or 5ch spine coil in MR1 if possible)

Plane Weighting Mode Slice Gap FAT SAT FOV Notes
SAG T2 DRIVE TSE 3mm 0.5mm None 20cm Include Facet Joints
SAG T1 TSE 3mm 0.5mm None 20cm Include Facet Joints
SAG STIR TSE 3mm 0.5mm STIR 20cm Include Facet Joints
SAG T2- FFE GRE 3mm 0.5mm None 20cm Include Facet Joints
AXIAL T2 FAT SAT TSE 3mm 0mm SPAIR 12cm Above cranio-cervical junction to Mid T1
AXIAL T1 TSE 3mm 0mm None 12cm Above cranio-cervical junction to Mid T1
COR T2 FATSAT TSE 3mm 0mm SPAIR 9cm Cranio-cervical junction