MR Pediatric Liver Mass WWO Protocol
Scan Notes:
- Images must be checked by radiologist or radiology resident before giving contrast and before the patient gets off the table.
- Make screenshot of amount of Volumen and glucagon was given and send to Pacs.
- Please do subtractions on mDixons: (subtraction = each of 3 post-contrast sequences – pre-contrast sequence).
Last updated:3/28/19
Charge as: Abdomen W/WO
Scanner preference: 1.5T or 3T
Coil: Torso or Cardiac
| Plane | Weighting | Mode | Slice | Gap | FAT SAT | FOV | Notes |
|---|---|---|---|---|---|---|---|
| AXIAL | T1 | TSE MV MS | 3.5mm | 1mm | None | Fit to Patient | Above diaphragm to iliac crests |
| AXIAL | T2 | TSE MV MS | 3.5mm | 1mm | SPAIR | Fit to Patient | Above diaphragm to iliac crests |
| COR | STIR | TSE MV MS | 3.5mm | 1mm | STIR | Fit to Patient | Entire abdomen, skin to skin |
| COR | DWIBS | IR EPI SS | 5mm | 1mm | None | Fit to Patient | Entire abdomen, skin to skin |
| Axial | 3D-mDixion | pre | 2.25eff | Fit to Patient | Scan Liver | ||
| COR | T1 | TSE MV MS | 3.5mm | 1mm | None | Fit to Patient | Optional: Only if requested by a radiologist.Entire abdomen, skin to skin |
| SAG | STIR | TSE MV MS | 3.5mm | 1mm | STIR | Fit to Patient | Optional: On any patient with neuroblastoma or paraspinal mass- or if requested by a radiologist.Entire abdomen, skin to skin |
Contrast injection
Power Inject: Bolus trak in sagittal plane: trigger when contrast reaches SMA
| Plane | Weighting | Mode | Slice | Gap | FAT SAT | FOV | Notes |
|---|---|---|---|---|---|---|---|
| AXIAL | 3 phase dynamic 3D-mDixon | post | 2.25eff | Fit to Patient | Bolus trak in sagittal plane: trigger when contrast reaches SMA: 3 phases, one right after another (roughly 30, 60, 90 seconds).Scan liver. | ||
| AXIAL | 3D-mDixon | delayed | 2.25eff | Fit to Patient | Exactly 4 minutes after contrast injection.Scan liver. | ||
| AXIAL | T1 | TSE MV MS | 3.5mm | 1mm | SPIR | Fit to Patient | Above diaphragm to iliac crests |
| COR | T1 | TSE MV MS | 3.5mm | 1mm | SPIR | Fit to Patient | Entire abdomen, skin to skin |
| AXIAL | 3D-mDixon | delayed | 2.25eff | Fit to Patient | Exactly 10 minutes after contrast injection.Scan liver. | ||
| SAG | T1 | TSE MV MS | 3.5mm | 1mm | SPIR | Fit to Patient | Optional: On any patient with neuroblastoma or paraspinal mass- or if specifically requested by a radiologist.Entire abdomen, skin to skin |