MR Neck WO - Sleep Apnea Protocol

Exam notes:

  • Exam should be proctored with radiologist present.
  • Use 3D PD to plan rest of scan. Respiratory trigger decreases blurring on PD. 
  • BTFE: Target temporal resolution of 300–400 ms/image over 30–35 s.
  • STIR: Radial acquisition decreases blurring.
  • Static and Cine sequences may need to be repeated at the level of suspected airway obstruction.

Other notes:

Last updated: 3/8/22
Charge as: Neck WO
Scanner preference: MR2 only.
Coil: NV or Spine

Plane Weighting Mode Slice (mm) Gap (mm) FAT SAT FOV (cm) MPR Notes
SAG PD 3D TSE 1.6 -0.8 no 20 AX, COR Centered on base of tongue. Respiratory triggering.
AXIAL cine BTFE 3D 5 0 no 20 no Centered on base of tongue or epiglottis
SAG cine BTFE 3D 5 0 no 20 no Centered on base of tongue
AXIAL cine BTFE 3D 5 0 no 20 no Centered on nasopharynx
AXIAL STIR 2D TSE 4 0.4 YES 22 no Mid nasal cavity to T1. Radial acquisiton.
SAG STIR 2D TSE 4 0.4 YES 22 no L-R coverage only of airway. Radial acquisiton.