MR Brain Vasculitis WWO Neuro Protocol
Coverage on VWI:
- If Philips: Oblique axial centered on COW, as in graphic below. F-H coverage 4.5-5.5 cm. Use MRA to plan placement of VWI sequence.
- If Siemens: Sagittal centered on COW. L-R coverage ~ 13.5 cm.
- Last updated: 9/26/2022
- Charge as: Brain WWO & MRA Brain WO
- Scanner preference: MR3, MR4, CHMR1. Only use 1.5T if implants.
- Coil: Highest channel available for max SNR (64 on Siemens).
| ?Plane | Weighting | Mode | Slice (mm) | Gap (mm) | FAT SAT | FOV (cm) | MPR | Notes |
|---|---|---|---|---|---|---|---|---|
| AXIAL | TOF | 3D FFE | 1 | 0 | no | 20 | MIP COW, Right, Left and Posterior | |
| AXIAL | DWI | EPI | 3 | 0.3 | YES | 23 | no | Angle to Corpus. Cover Skull Base to Vertex. Send only B1000 & ADC. |
| AXIAL | SWI | 3D GRE | 2 | -1 | no | 23 | no | Angle to Corpus. Cover skull base to vertex. Ok to add slices. Run Ven Bold on MR1 and MR2. |
| Siemens - SAG | T1-VWI | 3D TSE | 0.66 | 0 | YES | 21 x 21 | no | center on COW |
| Philips - AXIAL | T1-VWI | 3D TSE | 1.5 | -0.75 | no | 18 x 18 | no | center on COW |
| HAND INJECT CONTRAST | ||||||||
| SAG | FLAIR | 3D IR-TSE | 1 | 0 | YES | 23 | no | NO ANGLE. Cover ears and nose. Spacing and gap are variable. |
| Siemens - SAG | T1-VWI | 3D TSE | 0.66 | 0 | YES | 21 x 21 | SAG, COR | center on COW |
| Philips - AXIAL | T1-VWI | 3D TSE | 1.5 | -0.75 | no | 18 x 18 | SAG, COR | center on COW |
| SAG | T1 | 3D TSE | 1 | 0 | YES | 23 | AXIAL, COR | NO ANGLE. Cover ears and nose. Spacing and gap are variable. |