MR Wrist WO Protocol
Revised - 11/12/2018
- Charge as Wrist WO
- 3T Scanner Only
- Wrist/ Flex Coil
- Always Scan Individually
- Please ensure good signal with the smallest possible FOV
- If hardware is present:
- Do STIR instead of fat-sat mid-TE
- Do non-fat-sat T1 post-contrast
| Plane | Weighting | Mode | Slice | Gap | FAT SAT | FOV | Scan Range |
|---|---|---|---|---|---|---|---|
| AXIAL | PD (TE=20msec) | TSE | 2mm | 0.5mm | None | 12cm | 3cm Above/Below Wrist |
| AXIAL | Mid TE T2 Fat Sat | TSE | 2mm | 0.5mm | SPAIR | 12cm | 3cm Above/Below Wrist |
| COR | T1 | TSE | 2mm | 0.5mm | None | 12cm | Angle Relative to Radio-Ulnar Joint |
| COR | Mid TE T2 Fat Sat | TSE | 2mm | 0.5mm | SPAIR | 12cm | Angle Relative to Radio-Ulnar Joint |
| SAG | Mid TE T2 Fat Sat | TSE | 2mm | 0.5mm | SPAIR | 12cm | Angle Relative to Radio-Ulnar Joint |