MR Adult Brain Venous Thrombosis / MRV W/WO Protocol

Revised - 2/24/2011
Head Coil
Localizer
(REF scan)
Mode Slice Gap FAT SAT FOV Scan Range
SAG T1 TSE 4mm 1mm None 23cm Scalp to Scalp
AXIAL DWI EPI 3mm 1mm SPIR 23cm Angle to Corpus- Skull Base to Vertex
AXIAL T2 TSE 4mm 1mm None 23cm Angle to Corpus- Skull Base to Vertex
AXIAL T1 SE 4mm 1mm None 23cm Angle to Corpus- Skull Base to Vertex
AXIAL T2 GRE 3mm 1mm None 23cm Angle to Corpus- Skull Base to Vertex
COR MRV 2D TOF 2D TOF 2mm -0.5mm None 23cm Frontal to Occipital
AXIAL MRV 2D TOF 2D TOF 2mm -0.5mm None 20cm Skull Base to Vertex
Contrast Injection
COR Flair IR 4mm 1mm None 22cm Whole Brain
AXIAL T1 FS SE 4mm 1mm SPIR 23cm Angle to Corpus- Skull Base to Vertex
SAGITAL 3D T1 FFE None 23cm Scalp to Scalp
COR MRV 2D TOF 2D TOF 2mm -0.5mm SPIR 23cm Frontal to Occipital Bone
AXIAL MRV 2D TOF 2D TOF 2mm -0.5mm None 23cm Skull Base to Vertex