MR Adult Stroke Brain WO Protocol

Revised 2/17/2011
Charge for Brain WO
Head Coil
Localizer
(REF scan)
Mode Slice Gap FAT SAT FOV Scan Range
SAG T1 TSE 4mm 1mm None 23cm Scalp to Scalp
AXIAL T2 TSE 4mm 1mm None 23cm Angle to Corpus- Skull Base to Vertex
AXIAL T1 TSE 4mm 1mm None 23cm Angle to Corpus- Skull Base to Vertex
AXIAL T2* GRE GRE 4mm 1mm None 23cm Angle to Corpus- Skull Base to Vertex
AXIAL DWI 2mm Voxel SE EPI 3mm 0.3mm SPIR 23cm Angle to Corpus- Skull Base to Vertex
AXIAL FLAIR TSE 4mm 1mm None 23cm Angle to Corpus- Skull Base to Vertex
AXIAL ASL Single Phase FFE EPI 6mm 1mm None 23cm Place from top down. Do Not Send RAW Data to PACS
AXIAL ASL Multi Phase FFE EPI 6mm 1mm None 23cm See screen capture for slice placement. Archive RAW Data to DVD
MR Adult Brain WO
ASL Single Phase Slice positioning. Place slices from top down.
Peds Stroke Brain ASL Single Phase
ASL Multi Phase Positioning
Peds Stroke Brain ASL Multi Phase