MR Adult Frameless Brain W,WO, or W/WO plus DCE/DSC Perfusion Protocol

Revised - 3/23/2018

 
    POSITIONING THE PATIENT: Carefully Position Sponges Near Patients Chin. Cover at least 1 cm above the vertex of the brain - NO Angles. Cover the entire nose. ON 3D SEQUENCES, INCLUDE ADDITIONAL AIR ON TOP OF SCAN FOR SLICE WRAP

    POWER INJECTION:
    Prep Power Injector to Split Contrast Injection for DCE (1/3 of total contrast amount) and DSC (2/3). 15 ml Saline Chaser after DCE and DSC. PATIENT NEEDS A 20G OR LARGER PERIPHERAL IV.
8CH Head Coil
DCE Perfusion Sequence
(Dynamic Contrast Enhanced Perfusion)
Sequence Mode Dynamics FAT SAT FOV Coverage
AX 3D T1 FFE TSE 80 Dynamics None 22cm No Angle. Center through lesion of interest. Do not add slices!
DCE Notes:
  • Exam is set up with a manual start
  • Start injection @ 4-5 ml/second AT THE SAME TIME YOU BEGIN Dynamic 8
  • 15ml Saline chaser
DSC Perfusion Sequence
(Dynamic Susceptibility Contrast Perfusion)
Sequence Mode Dynamics FAT SAT FOV Coverage
AX 3D T1 FFE EPI 60 Dynamics None 20cm No angle. This should be set up already, based on DCE position. Do not change. Do not add slices.
DSC Notes:
  • Scan for 8-10 Seconds then begin contrast injection @ 5ml/s
  • 15ml Saline chaser
Stereotactic Frameless Sequences
Sequence Mode Slice Gap FAT SAT FOV Scan Range
AXIAL 3D T1 FFE 3D GRE 1.3mm 0mm None 25cm Cover the entire nose (A to P and S to I).1cm Below Skull Base to 1cm Above Vertex
AXIAL FLAIR 3D GRE 1.3mm 0mm None 25cm Cover the entire nose (A to P and S to I). 1cm Below Skull Base to 1cm Above Vertex
AXIAL T2 VISTA 3D GRE 1.5mm 0mm None 25cm Cover the entire nose (A to P and S to I). 1cm Below Skull Base to 1cm Above Vertex
Notes:
  • NO ANGLE, COVER ENTIRE BRAIN- ON 3D's INCLUDE ADDITIONAL AIR ON TOP OF SCAN FOR SLICE WRAP
PERFUSION POST PROCESSING
You Must perform the post processing for the perfusion.
Follow these instructions to complete the processing.
Brain Frameless