MR CSP WWO Protocol

Last updated: 10/16/2020
Charge as: CSP WWO
Scanner preference: 1.5T or 3T
Coil: NV Array Coils only whenever possible

Plane Weighting Mode Slice Gap FAT SAT FOV Notes
SAG T2 TSE 3mm 0.5mm None 20cm Lateral Sides of Vertebral Bodies
SAG T1 TSE 3mm 0.5mm None 20cm Lateral Sides of Vertebral Bodies
SAG STIR TSE 3mm 0.5mm STIR 20cm Lateral Sides of Vertebral Bodies
AXIAL T2 TSE 3mm 0.5mm None 12cm Above cranio-cervical junction to Mid T1
AXIAL T1 TSE 3mm 0.5mm None 12cm Above cranio-cervical junction to Mid T1

Contrast injection

Plane Weighting Mode Slice Gap FAT SAT FOV Notes
SAG T1 FATSAT TSE 3mm 0.5mm SPIR 20cm Lateral Sides of Vertebral Bodies
AXIAL T1 TSE 3mm 0.5mm None 12cm C2 DISC TO C7 DISC