MR CSP WO Protocol

Last updated: 10/16/20
Charge as: CSP WO
Scanner preference: 1.5T or 3T
Coil: 16ch NV with Anterior Element (do not use top-off in MR1 if possible)

Plane Weighting Mode Slice Gap FAT SAT FOV Notes
SAG T2 DRIVE TSE 3mm 0.5mm None 20cm Lateral Sides of Vertebral Bodies
SAG T1 TSE 3mm 0.5mm None 20cm Lateral Sides of Vertebral Bodies
SAG STIR TSE 3mm 0.5mm STIR 20cm Lateral Sides of Vertebral Bodies
AXIAL T2 TSE 3mm 0.5mm None 12cm Above cranio-cervical junction to Mid T1
AXIAL 3D T2 DRIVE TSE 2mm 0mm None 12cm Above cranio-cervical junction to Mid T1