MR Pediatric CSP WO Chiari 0-2yrs Protocol

Revised 5/23/2011

Charge as C-SPINE WO 

0 to 2 Years

NV Coil
Must include posterior fossa and all of 4th ventricle.
Localizer
(REF scan)
Mode Slice Gap FAT SAT FOV Scan Range
SAG T2 TSE 2.5mm 0.5mm None 20cm Include Facet Joints
SAG T1 TSE 2.5mm 0.5mm None 20cm Include Facet Joints
AXIAL T2 TSE 3mm 0mm None 15cm Skull Base to T1
COR THRIVE 3D FFE 1mm 0mm None 20cm Anterior Body to Spinous Process
Chiari CSP WO