MR Adult/ Pediatric Dynamic Pituitary W/WO Protocol
Revised - 2/14/2013
Charge as Brain W/WO
If the MRA is performed, charge for MRA Brain WO as well
8CH Head Coil
Localizer (REF scan) |
Mode | Slice | Gap | FAT SAT | FOV | Scan Range | |
---|---|---|---|---|---|---|---|
SAG | T1 | TSE | 2mm | 0mm | None | 15cm | Pituitary only |
COR | T1 | TSE | 2mm | 0mm | None | 13cm | Pituitary only |
COR | T2 | TSE | 2mm | 0mm | None | 13cm | Pituitary only |
CONTRAST INJECTION:
|
|||||||
COR | DYNAMICS | TSE | 3mm | 0.2mm | None | 15cm | Dynamic scans to cover the Pituitary only. Each Dyanamic should not exceed 20 seconds. |
SAG | T1 FAT SAT | TSE | 2mm | 0mm | SPIR | 15cm | Pituitary Only |
COR | T1 FAT SAT | TSE | 2mm | 0mm | SPIR | 13cm | Pituitary Only |